LI Guodong, JING Yachun, NIU Lina, WU Qiongye. Ultrasonic gas meter instantaneous flow measurement method and uncertainty evaluation[J]. Industrial Metrology, 2025, 35(4): 104-107. DOI: 10.13228/j.boyuan.issn1002-1183.2024.0279
Citation: LI Guodong, JING Yachun, NIU Lina, WU Qiongye. Ultrasonic gas meter instantaneous flow measurement method and uncertainty evaluation[J]. Industrial Metrology, 2025, 35(4): 104-107. DOI: 10.13228/j.boyuan.issn1002-1183.2024.0279

Ultrasonic gas meter instantaneous flow measurement method and uncertainty evaluation

  • Based on the instantaneous/cumulative flow method gas meter verification device developed by our unit, our hospital has become the first domestic measurement and testing institution that adopts instantaneous flow method to measure ultrasonic gas meters.This paper takes the ultrasonic gas meter compiled according to T/CMA-RQ “Optical Interface and Communication Protocol for Gas meter Detection”as the test object, introduces the measurement principle of infrared acquisition for velocity gas meter, and evaluates the uncertainty of the indication error measurement results measured by instantaneous flow method, aiming at providing reference for domestic relevant testing institutions.