医用体外引发碎石装置X射线定位系统检测方法

Testing method for X-ray positioning system of medical extracorporeal lithotripsy device

  • 摘要: 医用体外引发碎石定位装置X射线辐射源主要用于医疗单位泌尿碎石手术的检查和定位,其定位性能的准确性对碎石效果起着关键性作用,结合国家和行业标准及现场试验,确定辐射输出的空气比释动能率、辐射输出的质、空间分辨力、低对比度分辨力、辐射束相对于等中心的偏移及定位装置的最小位移为计量性能测试指标并提出计量检定方法,结果表明该方法科学可行,可操作性强,为碎石治疗设备的质量控制提供了有力的技术支撑和法规保障,填补了国内对碎石治疗设备无计量检定技术法规依据的空白。

     

    Abstract: The X-ray radiation source of the medical extracorporeal induced lithotripsy positioning device is mainly used for the inspection and positioning of urological lithotripsy surgery in medical units. The accuracy of its positioning performance plays a key role in the effectiveness of lithotripsy. Combining national and industry standards and on-site experiments, the air specific release energy rate of radiation output, the quality of radiation output, spatial resolution, low contrast resolution, the offset of radiation beam relative to the isocenter, and the minimum displacement of the positioning device are determined as the metrological performance testing indicators, and a metrological verification method is proposed. The results show that this method is scientifically feasible and highly operable, providing strong technical support and regulatory guarantee for the quality control of lithotripsy treatment equipment, filling the gap in China where there is no metrological verification technical regulatory basis for lithotripsy treatment equipment.

     

/