低能乳腺X射线辐射装置的研究与建立

Research and establishment of low-energy mammography X-ray radiation device

  • 摘要: 为解决医用乳腺X线机量值溯源问题,文章设计并建立了一套低能乳腺X射线辐射装置。该装置采用钼靶与钨靶双辐射源切换设计,可实现RQR-M和RQA标准辐射质的模拟;采用自上而下的垂直照射方式,真实模拟临床乳腺X射线机的工作状态,从而提升量值传递的有效性并减少散射线对周围环境的影响。本文详细阐述了辐射装置的整体结构、工作原理及辐射场的两种关键参数(同轴性与均匀性)测量方法,并重点对距X射线焦斑0.6 m处的辐射场均匀性进行了实验测试。结果表明,该位置均匀性大于95%的辐射野范围约为95 mm(水平)×90 mm(竖直),能够满足市售乳腺剂量仪电离室的校准需求。经过系统性能测试与长期实际应用验证,该装置运行稳定、可靠,为乳腺X射线诊断设备的量值溯源、辐射防护安全以及公众健康保障提供了重要的技术支撑。

     

    Abstract: To address the traceability challenges in medical mammography X-ray machines, this study designed and developed a low-energy mammography X-ray radiation device. The system features a dual-source configuration(molybdenum target and tungsten target) for generating RQR-M and RQA reference radiation qualities, and employs a top-down vertical irradiation geometry to accurately simulate the operational conditions of clinical mammography systems. This design enhances the efficiency of value transmission and minimizes the impact of scattered radiation on the surroundings. This paper elaborates on the core components, operational principles, and two measurement methodologies(coaxiality and uniformity) for characterizing the radiation field. Experimental focus was placed on the field uniformity at a distance of 0.6 m from the X-ray focal spot. Results demonstrate that the radiation field area with uniformity exceeding 95% is approximately 95 mm(horizontal) × 90 mm(vertical), which meets the calibration requirements for ionization chambers of commercial mammography dose meters. Repeated performance tests and long-term field application have confirmed the stability and reliability of the device, providing crucial technical support for the traceability of mammography equipment, radiation hazard control, and public health protection.

     

/

返回文章
返回