投审稿入口

酸溶氧化-电感耦合等离子体原子发射光谱法测定煤和矸石中镓钒磷

Determination of gallium,vanadium and phosphorus in coal and gangue by inductively coupled plasma atomic emission spectrometry with acid dissolution oxidation

  • 摘要: 准确测定煤和矸石中Ga、V、P元素含量,对伴生元素资源化利用、煤质评价及环境污染评估具有重要意义。现行国家标准方法(GB/T 8208—2007、GB/T 19226—2003和GB/T 216—2003)需在不同温度下高温灰化样品后采用分光光度法分别测定Ga、V、P,流程相对繁琐,且高温过程易导致部分元素挥发损失,无法实现多元素同步前处理与测定。本文基于HNO3、H2SO4和H2O2的强氧化性,将样品中有机质氧化为CO2予以去除,有效消除含碳物质对Ga、V、P的吸附与包裹作用,避免了高温灼烧除碳过程;此外,引入HF,与HNO3、H2SO4构成混酸体系,显著提升了含硅矿物的分解效率,据此实现了电感耦合等离子体原子发射光谱法(ICP-AES)对煤和矸石样品中Ga、V、P的同步测定。各元素校准曲线相关系数(r)均为1.000 0,Ga、V、P的检出限为0.29~2.43 μg/g,定量限为1.2~9.7 μg/g。按照实验方法测定煤和矸石中Ga、V、P,结果的相对标准偏差(RSD,n=7)为1.7%~9.4%;方法比对结果显示,本方法与国家标准方法的测定结果经t检验无显著性差异,由此证明所建方法准确可靠。

     

    Abstract: Accurate determination of gallium(Ga),vanadium(V) and phosphorus(P) contents in coal and gangue is of great significance for the resource utilization of associated elements,coal quality evaluation and environmental pollution assessment.Current national standard methods(GB/T 8208-2007,GB/T 19226-2003 and GB/T 216-2003) require high temperature ashing of samples at different temperatures,followed by separate determination of Ga,V and P via spectrophotometry.Such procedures are relatively cumbersome,and high temperature treatment easily causes volatile loss of partial elements,making simultaneous pretreatment and multi-element determination unavailable.In this study,based on the strong oxidation of HNO3,H2SO4,and H2O2,the organic matters in samples were oxidized and removed in the form of CO2,which effectively eliminated the adsorption and encapsulation of Ga,V and P by carbon-containing substances and avoided the carbon removal process by high temperature ignition.Furthermore,HF was introduced to form a mixed acid system together with HNO3 and H2SO4,which greatly improved the decomposition efficiency of silicon-bearing minerals.On this basis,a method for the simultaneous determination of Ga,V and P in coal and gangue sample by inductively coupled plasma atomic emission spectrometry (ICP-AES) was established.The correlation coefficients(r) of calibration curves for all elements were 1.000 0.The limits of detection(LOD) of Ga,V and P ranged from 0.29 μg/g to 2.43 μg/g,and the limits of quantification(LOQ) were between 1.2 μg/g and 9.7 μg/g.The contents of Ga,V and P in coal and gangue were determined according to the experimental method,and the relative standard deviations(RSD,n=7) of the determination results were between 1.7% and 9.4%.The method comparison results indicated that there was no significant difference between the proposed method and national standard methods via t-test, demonstrating the accuracy and reliability of the established method.

     

/

返回文章
返回